InTENSity CX4 Combination 4 Channel Electrotherapy & Ultrasound

Call for quote 800.461.0100


Login for pricing

SKU:
RM-DQ8000
Weight:
7.00 LBS
Health Canada Lic:
98881
CLEARANCE
InTENSity CX4 Combination 4 Channel Electrotherapy & Ultrasound

The InTENSity CX4 is an advanced combination electrotherapy and ultrasound system offering the practitioner a wide range of treatment options in one user-friendly, ergonomic design at a fraction of what other systems would cost. Equipped with a color, touchscreen menu-driven interface, the InTENSity CX4 intuitively groups and displays clinical protocols, guiding clinicians step-by-step to ensure the ideal therapeutic treatment. Parameters can be easily selected and adjusted. Clinicians can choose to use preset clinical protocols or customize their own. The comprehensive library provides recommended electrode placement to ensure better treatment outcomes and improved clinic efficiency. 

FEATURES & BENEFITS: 

Bilingual touchscreen user-friendly interface 

Dual frequency ultrasound at 1 MHz and 3 MHz 

Low BNR ratio of 5:1 

Multiple waveform electrotherapy: Interferential, Premodulated, TENS (Symmetrical Biphasic, Asymmetrical Biphasic, RAAS*), NMS, Russian, High Volt, and Microcurrent 

Rapid Agonist Antagonist Sequencing (RAAS*) replicates typical firing patterns of muscle contractions during functional movement for the best therapeutic outcome 

Independent intensity and parameter controls for each channel 

4 channels of independent stimulation - can be used in combination (linked) or totally independent 

Comprehensive full-color graphic anatomical and pathological libraries with recommended electrode placement 

"Favorites" feature stores up to 99 user-defined protocols for future use 

Sleek, user-friendly, ergonomic design for table top or therapy cart use 

 

BIPHASIC (TENS): 

Waveforms: 

Asymmetrical & Symmetrical Biphasic Rectangular 

Mode Selection: 

CC or CV 

Pulse Amplitude: 

CC: 0 – 200 mA 

CV: 0 – 200 V 

Pulse Width: 

20 – 400 μsec 

Pulse Frequency: 

1 – 250 Hz 

Treatment Time: 

1 – 60 minutes 

 

NEUROMUSCULAR STIMULATION (NMS) 

Mode Selection: 

CC or CV 

Pulse Amplitude: 

CC: 0 – 200 mA 

CV: 0 – 200 V 

Pulse Width: 

20 – 400 μsec 

Interphase Interval: 

120 μsec 

Pulse Frequency: 

1 – 200 Hz 

Ramp: 

1s, 2s, 5s 

Cycle Time: 

Continuous, 10/10, 10/20, 10/30, 10/50, Custom 

Treatment Time: 

1 – 60 minutes 

 

MICROCURRENT 

Frequency: 

0.1 - 1000 Hz 

Polarity: 

Positive, Negative, Alternating 

Pulse Amplitude: 

CC: 0 - 1,000 

Cycle (Fixed): 

50% 

Ramp (Fixed): 

1s 

Treatment Time: 

1 – 60 minutes 

 

IF-4P TRADITIONAL 

Waveform: 

Sinewave 

Mode Selection: 

CC or CV 

Vector Scan 

Auto: 20% – 100% 

Manual: 0° – 90° 

Carrier Frequency: 

Carrier F: 2 – 10 KHz 

Beat Low: 1 – (Beat H.) 

Beat High: (Beat L.) – 200 Hz 

Pulse Amplitude: 

CC: 0 – 200 mA 

CV: 0 – 200 V 

Treatment Time: 

1 – 60 minutes 

 

IFC-2P PREMODULATED 

Waveform: 

Sinewave 

Mode Selection: 

CC or CV 

Vector Scan: 

Auto: 20% – 100% 

Manual: 0° – 90° 

Carrier Frequency: 

Carrier F: 2 – 10 KHz 

Beat Low: 1 – (Beat H.) 

Beat High: (Beat L.) – 200 Hz 

Pulse Amplitude: 

CC: 0 – 200 mA 

CV: 0 – 200 V 

Ramp: 

2s 

Treatment Time: 

1 – 60 minutes

 

RUSSIAN 

Carrier Frequency: 

2.5 KHz 

Frequency: 

20 – 100 Hz 

Duty Cycle: 

10% – 50% 

Mode Selection: 

CC or CV 

Pulse Amplitude: 

CC: 0 – 100 mA 

CV: 0 – 100 V 

Ramp: 

1s, 2s, 5s 

Cycle Time: 

Continuous, 10/10, 10/20, 10/30, 10/50, Custom 

Treatment Time: 

1 – 60 minutes 

 

HIGH VOLT PULSED CURRENT (HVPC) 

Frequency: 

1 – 120 Hz 

Polarity: 

Positive or Negative 

Phase Duration: 

100 μsec 

Pulse Amplitude: 

CV: 0 – 500 V 

Ramp: 

1s, 2s, 5s 

Treatment Time: 

1 – 60 minutes 

 

 

CX4 ULTRASOUND PARAMETERS 

Frequency: 

1MHz ± 10% 

3MHz ± 10% 

Duty Cycle: 

10% – 100%, Continuous, Stepping 10% 

Pulse Duration: 

1 - 9 ms ± 10% 

(set by Duty Cycle) 

Pulse Frequency: 

100 Hz 

Ramp (Fixed): 

1s 

Treatment Time: 

1 - 60 minutes 

 

OUTPUT POWER 

Duty Factor: ≥80% for 5 cm2: 

0.5W – 10.0W 

Duty Factor ≤70% for 5 cm2: 

0.5W – 15.0W 

Duty Factor ≥80% for 1 cm2: 

0.1W – 2.0W 

Duty Factor ≤70% for 1 cm2: 

0.1W – 3.0W 

Output Accuracy 

± 20% (for any level above 10% of maximum) 

 

AMPLITUDE 

Duty Factor: ≥80%: 

2.0W/cm2 

Duty Factor ≤70%: 

3.0W/cm2 

Treatment Timer: 

0 – 30 min ± 0.1 min